b'Hyperion XTi Imaging System specificationsInstrument SpecificationsXTi Tissue Imager Dimensions* Description Specification Description SpecificationChannels 135 Dimensions Width: 58 cm (23 in)Height: 132 cm (52 in)Mass range 75209 amu Depth: 58 cm (23 in)Abundance sensitivity 0.3% 141Pr (M+1 for 140Ce)Weight 192 kg (423 lb)0.3% 174Hf (M1 for 175Lu) * The XTi Tissue Imager must be placed directly in front of theFrequency 800 pixels/sec CyTOF XT instrument.Detection limit 250 copies per cell Slide Loader (installed) Dynamic range 4.5 orders of magnitude Description SpecificationCalibration Automated Dimensions Width: 47 cm (19 in)Operating system Windows 10 Enterprise LTSC Height: 45 cm (18 in)Data storage 3.6 TB Depth: 19 cm (7 in)Cross-cell contamination2% (selected direction)Weight 20 kg (43 lb)(redeposition) Capacity 40 slides Crosstalk pixel to pixel 25%(20 slides per cassette)Maximum addressable16 x 58 mm Compatible slide L = 76 mm +0/1 mm (7576 mm)sample size dimensions W = 26 mm +0/1 mm (2526 mm)T = 1.1 mm +0.1/0.2 mm (0.91.2 mm)Brightfield field of view 250 x 250 m Note that beveled edge slides and chamber File type TXT, multipage TIFF, slides are not supported.OME-TIFF, OME-Zarr, MCD Supported barcodes QR, Data MatrixScan rate Cell Mode: 2 mm 2 /hr Minimum barcode5 x 5 mmTissue Mode: 50 mm 2 /hrdimensionsPreview Mode: 700 mm 2 /hrOptimized resolution Cell Mode: 1 m CyTOF XT Dimensions Tissue Mode: 5 m Description SpecificationPreview Mode: 1 m Dimensions Width: 93 cm (37 in)Lowest step size Cell Mode: 0.33 m Height: 135 cm (53 in)Tissue Mode: 5 m Depth: 75 cm (30 in)Preview Mode (fixed): 25 m Weight 288 kg (635 lb)Data Size File Type Size* MCD 45 MB/channelHyperion XTi 121701* 1,000 x 1,000 m region of interest (ROI) in Cell Mode Imaging System * 5,000 x 5,000 m ROI in Tissue ModeXTi Tissue Imager121000Learn more at Hyperion XTi 133068standardbio.com/xti Slide Loader 1519Email |[email protected] mWeb |standardbiotools-japan.com FLDM-01130 Rev 03 062025Hyperion XTi Unlocking New Views Brochure JPNFor Research Use Only. Not for use in diagnostic procedures.Patent and License Information: www.standardbio.com/legal/notices. Trademarks: www.standardbio.com/legal/trademarks. Any other trademarks are the sole property of their respective owners. 2025 Standard BioTools Inc. All rights reserved.'